In India, the Medical Devices Rules, 2017 (MDR 2017, notified as G.S.R. 78(E)) regulate every aspect of medical devices—from manufacturing and import to sale and post-market surveillance. One of the most critical yet often overlooked aspects is labelling. Proper labelling ensures patient safety, traceability, and regulatory compliance with the Central Drugs Standard Control Organization (CDSCO).
Labelling requirements are primarily detailed under Rule 44, with supporting provisions for exports (Rule 45), Unique Device Identification (Rule 46), shelf-life restrictions (Rule 47), and investigational devices (Rule 48). This comprehensive guide breaks down everything you need to know about medical device labelling in India.
1. Mandatory Labelling Requirements (Rule 44)
Every medical device sold in India must carry the following particulars in indelible ink on the label, shelf pack, and every outer covering:
|
Clause |
Requirement |
Key Points |
|
(a) |
Name of the medical device |
Exact notified name as per CDSCO classification |
|
(b) |
Details to identify the device & its use |
Intended use, user instructions (if needed) |
|
(c) |
Name & address of manufacturer |
Actual manufacturing premises (not just registered office) |
|
(d) |
Net quantity |
Weight, volume, number of units — only in metric system |
|
(e) |
Manufacture & Expiry date |
Format: “Expiry date MM/YYYY” or “Shelf Life” → Sterile devices: Date of sterilization can be treated as manufacture date → Stable materials (SS, titanium) or equipment: Expiry not mandatory |
|
(f) |
Medicinal/biological substance indication |
e.g., “Contains heparin”, “Drug-eluting stent” |
|
(g) |
Batch/Lot number |
Preceded by “Lot No.” / “Batch No.” / “B.No.” |
|
(h) |
Special storage/handling conditions |
e.g., “Store below 25°C”, “Protect from light” |
|
(i) |
Sterility status & method |
“Sterile”, “Sterilized by EO / Gamma” |
|
(j) |
Warnings / Precautions |
e.g., “Single use only”, “Latex-free” |
|
(k) |
Single use indication |
“Single Use” or ♻ with bar (ISO 7000-1051) |
|
(l) |
Physician sample |
“Physician’s Sample — Not to be sold” |
|
(m) |
Manufacturing Licence Number (Indian manufacturers) |
“Mfg. Lic. No. ML-XXX” |
|
(n) |
Import details (via sticker if not pre-printed) |
Import licence no., importer name/address, actual manufacturing site, manufacture date |
Symbols: BIS/ISO symbols can replace text only if meaning is obvious to the user.
Small-sized devices exemption: If label space is limited, only clauses (a), (b), (c), (d), (e), (g), (k), (m) are mandatory.
2. Labelling for Export (Rule 45)
Devices meant only for export can follow the importing country’s labelling laws, but minimum Indian requirements still apply on the innermost pack and all outer coverings:
Special cases (as requested by foreign buyer):
3. Unique Device Identification (UDI) – Rule 46
Mandatory from 1st January 2022
Every licensed medical device must bear UDI containing:
UDI must appear in both human-readable (HRI) and machine-readable (AIDC) formats (e.g., barcode or DataMatrix).
4. Shelf Life Restrictions (Rule 47)
|
Total Shelf Life Claim |
Minimum Residual Shelf Life Required at Import |
|
< 90 days |
≥ 40% |
|
90 days – 1 year |
≥ 50% |
|
> 1 year |
≥ 60% |
Example: A device with 24-month shelf life must have at least 14.4 months remaining at the time of import.
5. Labelling for Clinical Investigation / Demonstration (Rule 48)
Devices imported/manufactured for test, evaluation, clinical investigation, training, or demo need only the following on the container:
No need for full commercial labelling.
Common Non-Compliances Observed by CDSCO
Quick Checklist for Labelling Compliance (Rule 78E)
Conclusion
Labelling is not just a regulatory formality — it’s a patient safety tool. Non-compliant labelling is one of the top reasons for CDSCO detention, cancellation of licences, and import alerts.
Whether you’re a domestic manufacturer or an importer, Rule 78E compliance is non-negotiable.
Pro Tip: Use automated artwork management systems with UDI generators and pre-approved templates to avoid last-minute rejections.