The medical device regulatory framework in India is governed by the Medical Devices Rules (MDR), 2017, administered by the Central Drugs Standard Control Organization (CDSCO) under the Ministry of Health and Family Welfare.
Medical devices are classified into four categories according to their risk level: Class A (low risk), Class B (low-moderate risk), Class C (moderate-high risk), and Class D (high risk).
Within this framework, Class A non-sterile, non-measuring medical devices benefit from a simplified regulatory pathway. Under the amendment introduced through G.S.R. 777(E), dated 14 October 2022, these devices are exempt from specified licensing requirements. However, exemption from licensing does not mean exemption from all regulatory obligations. Mandatory registration, applicable labelling requirements, and other relevant compliance requirements remain important.
Class A non-sterile, non-measuring medical devices are medical devices categorized as low risk that are neither intended to be supplied in a sterile state nor designed to perform a measuring function.
To qualify for this category, a device must satisfy the applicable CDSCO classification criteria and fall within the relevant notified classification list.
The following characteristics help distinguish this category:
It is important to verify the classification of the specific device rather than relying solely on its commercial name or general description.
The primary regulatory provisions include:
CDSCO published an updated risk-based classification list for this category on 31 October 2025. Manufacturers and importers should consult the current list to confirm whether their particular device qualifies.
One of the most significant features of this regulatory category is the exemption from specified licensing requirements.
Under the applicable provisions, qualifying Class A non-sterile, non-measuring medical devices are exempt from the relevant licensing requirements under specified chapters of the Medical Devices Rules, 2017.
However, the exemption does not eliminate the need for mandatory registration.
Mandatory registration
Manufacturers and importers must furnish the prescribed information through the CDSCO Online System for Medical Devices and obtain a system-generated registration number.
According to CDSCO guidance:
Companies should verify the current online registration procedure and any applicable amendments before placing a device on the Indian market.
Manufacturers intending to market qualifying Class A non-sterile, non-measuring medical devices in India should prepare the prescribed registration information.
The information generally includes:
The exact information and declarations should be checked against the current requirements under the Medical Devices Rules, including the provisions governing this category.
Importers must also complete the prescribed registration process for qualifying Class A non-sterile, non-measuring devices.
The importer should verify the device classification, identify the manufacturer and manufacturing site, and ensure that the required product information and declarations are available.
Key compliance activities include:
The exemption should not be interpreted as a blanket exemption for all imported medical devices. Devices supplied sterile or having a measuring function may fall outside this category.
A common misconception is that exemption from licensing automatically removes quality management obligations.
This is not correct.
CDSCO guidance clarifies that the manufacturing facility for Class A non-sterile, non-measuring medical devices must conform to the applicable quality management system requirements for medical devices.
Manufacturers should establish and maintain suitable documented procedures covering areas such as:
The scope and implementation of the quality management system should be determined with reference to the applicable regulatory provisions and the device's characteristics.
Labelling is a key compliance requirement even when a device is exempt from specified licensing provisions.
The product label should comply with the applicable provisions of the Medical Devices Rules, 2017, and any relevant amendments or standards.
Depending on the applicable requirements, labelling information may include:
Manufacturers and importers should verify the precise labelling requirements applicable to their device instead of treating this list as a complete or universally applicable label template.
Labelling must also remain consistent with the registered product information and the manufacturer's intended use.
Manufacturers should ensure that the device conforms to the applicable essential principles of safety and performance.
This involves identifying relevant safety and performance requirements and maintaining appropriate evidence that the device meets them.
Depending on the device, supporting documentation may include:
Self-certification does not remove the manufacturer's responsibility to ensure that the device is safe and performs as intended.
Before applying the simplified registration pathway, manufacturers and importers should carry out a classification assessment.
A practical assessment should cover the following questions:
Step 1: Confirm the intended medical purpose.
Establish whether the product meets the applicable definition of a medical device under Indian law.
Step 2: Verify the risk classification.
Check the applicable CDSCO classification list and relevant classification rules.
Step 3: Confirm that the device is non-sterile.
A device intended to be supplied in a sterile state does not qualify for this category.
Step 4: Confirm that the device is non-measuring.
Assess whether the device is intended to measure a quantitative physiological or anatomical parameter, or another quantity covered by the applicable definition.
Step 5: Complete registration and compliance activities.
Submit the required information, obtain the registration number, and meet the applicable quality, labelling, safety and performance requirements.
If the device does not satisfy all relevant conditions, the manufacturer or importer should determine the appropriate regulatory pathway before marketing it.
Companies should be careful to avoid the following errors:
A documented regulatory assessment can help companies avoid these errors and support a smoother registration process.
The regulatory pathway for Class A non-sterile, non-measuring medical devices in India is designed to reduce specified licensing burdens while retaining important safeguards for product safety and performance.
Manufacturers and importers should focus on four essential activities: confirming the correct classification, completing mandatory registration, maintaining the applicable quality management system, and complying with labelling and other relevant requirements.
The latest CDSCO classification list and applicable amendments should always be reviewed before a device is registered or marketed. A device that falls outside the exemption must follow its applicable regulatory pathway under the Medical Devices Rules, 2017.
Last Updated : 14 Oct 2026